Reducing Unnecessary Repeat CT Examinations Through Imaging Protocol Optimization: A Retrospective Quality Improvement Study

Main Article Content

Sharifa Mohammed Ali Hadi, Abdulwahab Matar Awaad Almutairi, Ali Mohammed Ali Alshehri, Afaf Ali Hassan Shar, Sami Abdulrahman Fardis Alqarni, Noura Abdulhameed Omar Ayoob, Hajar Mohammed Hasan Altharwi, Amal Abdulrahman Ibrahim Alsabhani, Sara Omar Saleh Alsaiary, Ohud Abdullelah Mohammed Alkhudairi

Abstract

Background. Repeat computed tomography (CT) due to failure of the ordering process, technique or specific protocol, rather than in response to a change in the clinical question adds to the levels of radiation and healthcare costs, and increases time delays without further diagnostics, and is sparsely collected as a quality measure.
Objectives. To explore if a multicomponent CT protocol optimization bundle decreased the rate of unnecessary repeated CT examinations at a tertiary teaching hospital. Methods. Retrospective analysis of quality improvement with interrupted time series using three CT scanners (12-month pre-implementation, 1-month transition, 12-month post-implementation; 38,719 CT examinations). The bundle of protocol standardization, indication-based protocol assignment, technologist pre-scan check list, standardized bolus-tracking criteria, duplicate-order alert, and monthly audit-and-feedback. Two radiologists established re-examinations (k = 0.84). The primary outcome was unnecessary repeat CT per 1,000 examinations and was analyzed with quasi-Poisson segmented regression and a statistical process control chart. Secondary outcomes were cause-specific rates, dose, image quality, workflow, and costs; all clinically indicated re-examinations were a negative-control outcome.
Results. Unnecessary repeat CT decreased from 35.4 to 23.5 per 1,000 examinations (rate ratio 0.66 (0.59-0.75)). Segmented regression detected an immediate change in level of 0.74 (0.57-0.95) (p =.021) and a decreasing trend; an estimated 326 fewer repeat examinations (41.3% of the counterfactual) and 2,310 person-mSv were avoided. Largest reductions occurred in duplicate orders, protocol selection, and contrast-timing errors; contrast-absent examination results was unchanged. Median effective dose per examination declined from 5.8 to 5.1 mSv; image-quality scores did not decline; repeat examinations with a clinical indication (a negative control) was unaffected.
Conclusions. A simple, process-focused protocol optimization bundle was associated with a sustained, clinically significant reduction of unwarranted repeat CTs without a loss of image quality or throughput. Protocol optimization for motion repeats requires a different approach. A prospective multicentre trial is indicated.